6.5.15 Instrument ID, unique instrument number for identification
The device is given a unique ID with up to 4 digits in this menu.
This ID can be used when downloading the recorded data. This
enables the data to be assigned to a specific device or location.
1.
Call up CONFIG mode by pressing the UP or DOWN keys
until the arrow lights up beside CONFIG. Then press EXIT.
A3447
Fig. 31: ID
2.
Press the
[right arrow key] to move to the next menu until ID
is displayed.
3.
Use the UP or DOWN keys to select the value of the first
number.
Use the
[right arrow key] to select the next number and
repeat the process for all numbers.
Scroll through the numbers to the next menu.
6.5.16 Protocol interval, setting the interval for logged files
This menu defines how often the measured values are logged. The
setting is in minutes. The interval can be set at from 1 ... 60
minutes.
1.
Call up CONFIG mode by pressing the UP or DOWN keys
until the arrow lights up beside CONFIG. Then press EXIT.
A3448
Fig. 32: INVL
2.
Press the
[right arrow key] to move to the next menu until
INVL is displayed.
3.
Use the UP or DOWN keys to set the value.
Setting the device ID:
Setting the logging interval:
Commissioning
44