6.5.7 Response time
A0533
Fig. 23: Response time
Function
Options
Information
Response time (RESP) 1 ... 60 seconds
Factory setting: 10 sec‐
onds
Select a response time for the NTU values displayed
and output.
Select the highest response time (i.e. the highest
number) to avoid anomalies caused by air or other
influences.
Select the lowest response time (i.e. the lowest
number) if rapid changes are expected that need to
be monitored.
The displayed number is a relative response time.
The approximate response time (in seconds) can be
calculated by the number displayed multiplied by 5.
6.5.8 Display resolution
A0534
Fig. 24: Display resolution
Function
Options
Information
Resolution (RES)
1 ... 0.0001
Factory setting: 0.01
The DULCO turb C is able to dis‐
play values with a resolution of up
to 4 digits after the decimal point
with display values below 10 NTU