Function
Options
Information
Error Level(ERLV)
OFF
0 mA
2 mA
4 mA
6.5.3 Setting the Baud rate
A3416
Fig. 17: Setting the Baud rate
Function
Options
Information
485
Baud
Baud value
n
1200
n
2400
n
4800
n
9600
n
19200
Selection of the Baud rate for the I/O connector
Addr
1 ... 255
Selection of the device address
MBUS
ASCII
RTU
Selection of ASCII or RTU
6.5.4 Inserting the Modbus address – Selection of the address
Select the Modbus address of the DULCO turb C from 1 ... 255.
Additional information about Modbus can be obtained from ProMi‐
nent on request.
1.
Call up CONFIG mode by pressing the UP or DOWN key
until the arrow lights up beside CONFIG and then press the
EXIT key.
2.
Press the
[right arrow key] to move to the next menu until
ADDR is displayed.
How to change the Modbus address:
Commissioning
35