3.2 Specifications
Measuring range:
0 ... 100 NTU, factory-configured type TUC 5 (infra-red) & type TUC 6 (white
light) 0 ... 10 NTU & 0 ... 1000 NTU optional ranges (loaded on a USB
memory stick)
Precision:
± 2 % of the measured value or ± 0.02 NTU below 40 NTU, depending on
which value is greater (ranges 0 ... 100 and 0 ... 1000) ±5 % of the measured
value above 40 NTU (ranges 0 ... 100 and 0 ... 1000) ±2 % of the measured
value or ±0.02 NTU (range 0 ... 10 NTU)
Resolution:
0.0001 NTU (below 10 NTU)
Response time:
Can be set
Display:
Multi-row liquid crystal display with backlighting
Alarms:
Two programmable 120 ... 240 V AC, 2 A, form C relays
Analogue output:
4 ... 20 mA, 1000 Ω
Communication connector: Bidirectional RS-485, Modbus RTU
Data storage:
16 GB download using a USB memory stick
Software:
Can be upgraded using a USB memory stick
Maximum water pressure:
Integrated pressure regulating valve with 7 bar (101.5 PSI)
Flow rate:
100 ml/min ... 1.5 l/min (.026 ... .40 gal (US)/min)
Operating temperature:
1 °C ... 50 °C (34 °F ... 122 °F)
Wetted materials:
nylon, borosilicate glass, silicone, polypropylene, stainless steel, Viton, acetyl
Sample temperature:
1 °C ... 50 °C (34 °F - 122 °F)
Power supply:
100 ... 240 V AC, 47 ... 63 Hz, 80 VA
Insulation value:
Dual insulation, degree of contamination 2, overvoltage category II
Ambient conditions:
Not recommended for use outdoors. Max. operating altitude up to 2000
metres. Up to 95 % relative humidity (non-condensing).
Housing:
ABS
Compliance with statutory
regulations and certifica‐
tions:
White light version conforms to U.S. EPA 180.1
Infra-red version conforms to ISO 7027 and DIN 27027, CE-tested,
LC-listed in compliance with UL 61010-1, 3rd edition: 2012 LC: Certified to
CSA 22.2 No.61010-1-12: 2012
Instrument weight:
1.8 kg (4.0 lbs.)
Shipping weight:
2.8 kg (6.2 lbs.)
3.3 Ultrasonic cleaning
Ultrasonic cleaning is used for the continuous cleaning of the cuv‐
ette. It is not intended for cleaning cuvettes that are already dirty,
nor can it replace manual cleaning. However, the cleaning intervals
for manual cleaning can be significantly extended by the use of this
system. The use of a cuvette with an ultrasonic transducer is
required for operation. The system emits an ultrasonic frequency
via a spring to a piezoelectric transducer, which is connected to the
base of the cuvette through which the fluid flows. The cuvette's
detection system only operates in
[AUTO] operating mode.
Functional description / Product identification