Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516
5
www.promega.com
TM397 · Revised 12/16
UV-Visible Absorbance Module Specifications
Detector
Top-reading, head-on photon counting
photomultiplier tube (PMT)
Light Source
Xenon flash lamp
Spectral Range
200–600nm
Filter Wheel
9 absorbance filters included
Wavelengths Included
230, 260, 280, 320, 405, 450, 490, 560 and 600nm
10nm bandwidth
Detection Limit
0.1 O.D.
Dynamic Range
0–4.0 O.D. at 260nm and 280nm
Injector System Specifications
Number of Injectors (optional)
Two injectors
Dispense Volume Range
5–200µl in 1µl increments
Plate Compatibility
6-, 12-, 24-, 48- and 96-well plates
Injection Speed
20–500µl per second
Waste Collection Tray volume
Approximately 50ml
Void Volume
500µl
Shaker System Specifications
Shaking Pattern
Linear or orbital patterns
Shaking Intensity
100–500 cycles per minute
Shaking Orbit
1mm, 2mm and 3mm settings
Heating System Specifications
Temperature Range
A5°C to 45°C
Computer Requirements
Operating System
Windows
®
8.1 or 10, 64 bit, with USB connection
to instrument
RAM
2GB (4GB preferred)
Disk Space
64GB
Other Requirements
Microsoft.NET Framework 4.0 (full version),
Microsoft
Excel 2013 Professional