g. Selecting any well or set of wells will activate the well labeling options. The available types of well
information are curve fit-specific. For a Linear Standard Curve,
Unknown
,
Known
and
Blank
well
designations are available, while for Linear Response and Dose Response curve types, only Known and
Blank well designations are available. The well types correspond to:
Unknown:
A well where the concentration of the sample is unknown. It is possible to set a name and
dilution factor for this type of well. Unknown wells are only allowed in a Linear Standard Curve type
analysis template.
Known:
A well where the concentration of the sample is known. You can set a name and concentration
for this type of well.
Blank:
A well containing a blank for the experimental data set. The average values of all blank wells will
be subtracted from all Known and Unknown wells prior to analysis.
h. Begin creating the template by touching and dragging to select all wells that will contain samples of a
given type, and then touch the corresponding well type (
Unknown
,
Known
,
Blank
) that you wish to
assign to those wells (Figure 50).
To select nonadjacent wells, touch a well and hold down until a rectangle appears; then continue with
selection.
Fi
13908T
A
gure 50. Template with well designations assigned.
In this figure unknown, known and blank well
designations have been assigned across a plate.