TROUBLESHOOTING
MULTI GAUGE CONTROLLER
continued from previous page
NOTIFICATION
DESCRIPTION
High Pressure
Pressure in the gauge area is out of the
measuring range.
Check the gauge and its type whether it is appro-
priate for the application/process.
EEPROM Failure
Internal hardware error
Turn off, wait for
s, restart
Filament Failure
No current in the ilament circuit
Check whether the gauge ilament is burnt out
Check the connections of the controller
Replace the measurement cable
Restart emission
Sensor Break
Gauge damaged or no sensor
Check the connection of the gauge with the device
Check and replace the cable if necessary
Table . : Noti ications in pressure value/information box.
. MESSAGE BAR NOTIFICATIONS
. .
ERRORS
•
E : Connection with the mainboard lost Error code: <number>
–
Connection with the mainboard lost. The error code is displayed to ind the source of
the problem.
•
E : We recommend to delete unnecessary video iles.
–
The error caused by lack of free space on the SDHC card. To free space, remove several
videos from the card.
. .
. . WARNINGS
•
W :Low Disk Space: We recommend to remove videos from device.
We recommend
to remove videos from device. Warning resulting from insuf icient space on the SDHC card.
Delete videos in order to free up space.
.
. NOTIFICATIONS IN MESSAGE BOX
•
M : Copy summary: Error when copying a log ile.
Copying Log ile failed.
Page
User manual rev.