OPERATION
MULTI GAUGE CONTROLLER
. .
FILTER
Setup menu →Filter
Dependent on whether the user needs stable readings or prompt fast system response, he can
choose between three iltering options depending:
•
Low
- no averaging
•
Medium
- averaging out of samples
•
Heigh
- averaging out of
samples
. .
CHANGING NUMBER OF DISPLAYED CHANNELS - DISPLAYED CHANNELS MENU
Setup menu →Displayed channels
The menu enables the user to select the number of channels to be displayed on the main window.
The menu can display from to channels. The vacuum channels to which no gauge is connected
can be hidden.
. .
CHANGING CHANNEL NAMES - CHANNEL NAMES MENU
Setup menu →Channels Names
The menu enables the user to change the names of any channel.
The entered name appears on the main screen and in setpoint settings and all items in the setup
menu.
The length of a name is limited to characters.
. .
OUTPUTS SETTINGS
Setup menu →Outputs settings
. . .
RELAY OUTPUTS CONFIGURATION - RELAY OUTPUTS MENU
Setup menu →Outputs settings →Relay Outputs
The MG has relay outputs, which functioning may be bound with the current values of the pa-
rameters measured by the MG . The outputs are physically separated from the device (potential-
free). The outputs are available on the -pin I/O connection on the rear panel. The relay outputs
may be con igured by selecting outputs from
Output
to
Output
and assigning a required output
signal.
The following signal sources are available:
•
Emission
- the state of the output depends on the emission activation state
•
Degas
- the state of the output depends on degassing activation
•
Still OFF
- output inactive
•
Still ON
- output active
•
None