MULTI GAUGE CONTROLLER
INSTALLATION
INSTALLATION
This chapter describes the procedures for unpacking, mechanical installation and electrical instal-
lation. Take care when lifting the unit that the weight and position do not exceed comfortable limits.
. UNPACKING
. Visually inspect the transport packaging for signs of external damage.
. Unpack the MG .
Note: Retain the packaging materials for later use. The MG must be stored and trans-
ported in the original packaging material only.
. Examine the MG for completeness.
. Visually inspect the MG for signs of damage.
Damaged product.
Using the destroyed or damaged device or connection cables can result in elec-
tric shock. Never attempt to put a damaged product into operation. Secure
the damaged product from unintended operation. Send a damage report to the
courier company or the insurer.
. MECHANICAL INSTALLATION
The MG
device can be mounted in a DIN
( ”, HU) cabinet/rack, occupying half of its
width. In each of these cases you must pay attention to the following safety note.
Ambient temperature.
Exceeding the maximum permitted ambient temperature may damage the de-
vice.
Before switching on the device, make sure that the maximum permitted ambi-
ent temperature is not exceeded and that the air can low freely through the
vent holes.
Do not expose the device to direct sun.
User manual rev.