MULTI GAUGE CONTROLLER
OPERATION
•
Gateway
- default gateway on a TCP/IP network,
•
DHCP
- Dynamic ”Host” Con iguration:
–
enable
- automatic con iguration on connection (IP address, gateway, subnet mask),
–
disable
-IP protocol parameters must be set manually,
•
TCP server port
- speci ies the port number used in network socket created by TCP server,
default -
To change the
Interface
parameters enter
Setup menu →Device Settings →Communication
→Interface
.
The detailed description of the RS
/
connection pins is in: . .
•
Interface
- interface type:
–
RS
–
RS
–
Pro iNet (optionally)
•
Baud Rate
:
–
available Baud Rates:
bps,
bps,
bps,
bps,
bps,
bps,
bps,
–
bits of data, stop bit and no parity parameters are ixed and cannot be changed,
In the
Communication menu
, apart from the physical parameters of the communication interfaces,
the following parameters can be con igured:
•
Remote Control
- the option enables activation of the remote control mode. In the remote
control mode changing the parameters through the touch panel is locked. It is only possible
to view the parameters on the screen and view the device con iguration. The device in the
remote control mode displays in the upper part of the main window an information bar with
text: ”REMOTE: MG ”.
Activating the option does not mean that the device immediately switches to the remote con-
trol mode. It happens only after establishing communication with a remote control device.
Then the edition of the Remote Control option is locked and local control via the touch panel
is not possible. The remote control can be disabled only in two cases:
–
failure of communication for more than s,
–
sending a special command resulting in entering into the local control mode.
Entering the
Remote Control
mode is necessary if the device must be controlled through a
communication interface. Reading of the parameters, however, is always possible and does
not require the mode to be activated.
•
Address
- current communication address of the device. Depending on the type of the device
this item can be edited, or can display the address that has been set on the DIP switch located
on the rear panel of the device.
User manual rev.
Page