TROUBLESHOOTING
MULTI GAUGE CONTROLLER
TROUBLESHOOTING
This section describes the errors, warnings and noti ications appearing in the MG . All informa-
tion displayed on the screens is divided into two groups:
• Errors and warnings , which relate primarily to the device hardware are displayed in the mes-
sage bar on the right side of the screen. They are identi ied by the colours red and yellow
(more details about them can be found in the chapter on user interface). An error disappears
from the bar when it will be physically eliminated.
• All information that is not directly related to operation of the device is displayed as a message.
This information presents noti ications after user interaction. More information can be found
in chapter . . .
. NOTIFICATIONS IN PRESSURE VALUE/INBFORMATION BOX
NOTIFICATION
DESCRIPTION
STATUS
Wait for emission
Waiting for starting the gauge emission
Wait: the device should automatically start the
gauge
Check the type of the selected gauge
Check the measurement cable/replace the mea-
surement cable
Emission off
Emission is turned off
To turn on emission press the Emiss button
Emiss
Degasing [time]
Gauge degassing in progress
Wait until the device inishes degassing
Offset Calib.
The device is in the internal calibrating
mode
Wait until the device inishes calibration
ERRORS
ShortFilament
Current in the ilament circuit is too
high, and emission current cannot be
achieved
Check the ilament for short-circuit with ground
or other contacts