58
Liquidus Tracker
4. If you do abort the run, the controller will stop the profile and begin to write the data to the USB
flash drive.
Writing to USB disk - DO NOT SWITCH
OFF
Progress 025%
Note
Switching the system off during a write to the USB flash drive may result in a permanently corrupt
disk.
5. The data are stored to the USB flash drive as a pdf file with the prefix PIDA. See
Writing the last
run to a USB flash drive
.
6. Once the data has been written the system will move to the warm-up cycle.
3.4.5
Alarm signals
During the run, the controller checks the current temperature against the programmed demand.
Should the chamber deviate by more than 10 °C from the required profile, the
information LED
will
flash and the internal sounder will produce a continuous alarm tone. In addition the alarm output
contacts will open. These may be used to trigger an external alarm.
Pressing any key will silence the alarm without affecting any other aspects of the run.
3.5
Profile warm-up cycle
At the end of a run due to the run being aborted or reaching its normal conclusion, the system will
check to see if the chamber is warm. If it is colder than 10 °C you will offered the chance to run a
warm-up cycle.
Return chamber to ambient? : Run
aborted
<Run warm-up profile> CANCEL
Use the cursor keys to select either
Run warm-up profile
or
CANCEL
.
When you select the
warm-up
profile, a new profile run will begin to take the chamber to 20 °C.
Once the temperature has been reached the system will return automatically to the main menu.
59
10
Содержание Kryo360-1.7
Страница 1: ...Liquidus Tracker Operator s Guide...
Страница 5: ...Introduction Section I...
Страница 20: ......
Страница 21: ...Installing the system Section II...
Страница 29: ...25 Installing the system...
Страница 33: ...29 Installing the system 10 Connect the power supply cables to the pumps and chamber...
Страница 35: ...Operating the system Section III...
Страница 65: ...Routine maintenance and troubleshooting Section IV...
Страница 68: ......
Страница 69: ...Additional information Section V...
Страница 76: ......
Страница 80: ...MA101894 Issue 2 Liquidus Tracker Operator s Guide...