Liquidus Tracker
7. Fit the locating plate on top of the mixing blades.
8. Close the extraction line clamp
(5)
.
9. Fill the pot with the starting volume of cryoprotectant. This should be level with the top of the
extraction tube in the carrier and is the volume defined as
Vcon
in the liquidus tracker
configuration. See
Liquidus tracking parameters - single solution mode
. The concentration
must also be the same as the
Kstart
parameter.
10.Ensure that the cryoprotectant covers the top surface of the condyle. If it does not, either
packing pieces should be added to displace the solution or more cryoprotectant will need to be
added and the Liquidus tracking parameter readjusted.
11.Screw the lid of the pot onto the carrier.
12.Connect the injection line
(10)
to the container containing the cryoprotectant
(1)
. Unless a
collapsible container such as a flexible bag is used, this container will need to be fitting with an
inlet filter
(9)
to maintain sterile conditions in use.
13.Connect the extraction line
(11)
to the waste collection container
(6)
. Unless an expandable
container such as a flexible bag is used, this container will need to be fitting with an exhaust filter
(8)
to relief any pressure build up whilst maintaining sterile conditions in use.
14.Connect the pressure relief line
(7)
to the waste collection container
(6)
.
40
Содержание Kryo360-1.7
Страница 1: ...Liquidus Tracker Operator s Guide...
Страница 5: ...Introduction Section I...
Страница 20: ......
Страница 21: ...Installing the system Section II...
Страница 29: ...25 Installing the system...
Страница 33: ...29 Installing the system 10 Connect the power supply cables to the pumps and chamber...
Страница 35: ...Operating the system Section III...
Страница 65: ...Routine maintenance and troubleshooting Section IV...
Страница 68: ......
Страница 69: ...Additional information Section V...
Страница 76: ......
Страница 80: ...MA101894 Issue 2 Liquidus Tracker Operator s Guide...