Liquidus Tracker
1. DMSO solution. Concentration must match the configuration setting.
2. Peristaltic pump 1
3. 0% solution
4. Peristaltic pump 2
5. Y piece
6. Sample container
7. Waste collection container
8. Filtered exhaust vent
9. Filtered inlet vents
10.Injection line 1
11.Injection line 2
12.Combined injection line
13.Extraction line
14.Pressure relief valve
Warning
Ensure that a pressure relief valve has been fitted between outlet of pump 1
(2)
and the waste
collection container
(7)
. This is to ensure that in the event of the cryoprotectant freezing, the
pumps will not overpressurise the system and cause a leak of fluid. The pressure relief valve
should be set to less than the rated tubing pressure and no more than 1.24 bar ( 18 psi).