Operating the system
15.The completed assembly, illustrated in the schematic diagram below can now be taken to the
liquidus tracker chamber.
16.Slide the condyle container into the main stirring lid. Rotate the shaft of the stirrer if necessary to
get the magnetic coupling to align.
17.Push the three lines from the container (pressure relief line
(7)
, injection line
(10)
and extraction
line
(11)
, into the grooves in the underside of the main stirring lid so that they are retained in the
grooves.
18.Lower the lid into the main chamber.