41
Operating the system
longer than the injection stage to allow the pump more time to operate in suction mode.
This cycle then repeats as shown below to produce an injection-stand-extraction cycle.
Tip
Make sure the
Single solution
check box is ticked in the
Special values for PID Extended
window
Vcon
defines the volume of cryoprotectant in the sample container at the start of the run.
F1
,
F2
,
Off1
and
Off2
are the pump flow factors that define the flow rate for each pump in mL/rpm.
Flow rate = Fn * rpm + Offn
Ksol1
is the concentration of the solution delivered by pump 1. The solution provided by pump 2 is
assumed to have a concentration of 0%.
Setting
F1
,
F2
or
Ksol1
to 0 will disable the pumps.
KStart
is the concentration of cryoprotectant that is in the sample carrier at the start of the run.
The remaining parameters are associated with the temperature-concentration curve. See
Temperature-concentration table
.