LIFEPAK 20/20e Defibrillator/Monitor
Modes of Operation
Inservice Mode
Introduction
Inservice mode enables users to practice or demonstrate the monitoring
functions of the device. The functions include:
■
Selecting ECG lead selection, size, and volume, and moving ECG waveform
with heart rate
■
SpO2
■
Alarms
■
Events
Note:
No therapy features are available in the Inservice mode.
Entering Inservice
Mode
To enter Inservice mode:
1. Remove all cables from the device. Inservice mode cannot be entered if
cables are attached to the device.
2. While holding down the
HOME
and
EVENT
buttons, turn the device ON.
Release these buttons when the
INSERVICE
overlay appears.
To exit Inservice mode, turn the device OFF.
4-16