LIFEPAK 20/20e Defibrillator/Monitor
Troubleshooting
Service Code Table
(continued)
(Continued on next page)
Code
Service Code Description
Corrective Action
9c29
3:00 AM redundant controls test: redundant controls
stuck on
Replace the A04 Therapy PCB
9c2a
3:00 AM redundant controls test: enable 2 stuck on
Replace the A04 Therapy PCB
9c2b
3:00 AM redundant controls test: enable 1 stuck on
Replace the A04 Therapy PCB
9c2c
Capacitor voltage per pacing pulse too high
Replace the A04 Therapy PCB
9c2d
Capacitor current per pacing pulse too high
Replace the A04 Therapy PCB
9c2e
Cap current per pacing pulse too high
Replace the A04 Therapy PCB
9c2f
Pacing current and selected current out of range
Replace the A04 Therapy PCB
9c30
Pacing pulse width too short
Replace the A04 Therapy PCB
9c31
Pacing pulse width too long
Replace the A04 Therapy PCB
9c32
Capacitor voltage and predicted capacitor voltage
mismatch
1. Energy capacitor is not connected or is open
2. Replace the A04 Therapy PCB
9c35
Therapy CPU instruction test failed
Replace the A04 Therapy PCB
6-38