LIFEPAK 20/20e Defibrillator/Monitor
Replacement Procedures
Boardstack
(continued)
Replacement Procedure A - Keystone:
Caution: The positive contacts of the Keystone coin cell battery holder
are easily damaged. Do not bend or deform the positive contacts during
removal or insertion. If the battery holder is damaged, replacement of the
A01 System Controller PCB is required.
1. Push battery gently toward the positive side of battery holder while using a
flat screwdriver to lift up the edge of the battery closest to the negative side
of the holder (opposite the positive contacts) until the battery releases from
the holder.
9-92