LIFEPAK 20/20e Defibrillator/Monitor
Replacement Procedures
Boardstack
(continued)
A02 Patient Parameter
and A06 OEM/SpO2
Assembly Installation
(continued)
b. Install the Nomex shield (262) onto the OEM PCB by securing it with two
washers (264) and nuts (161); torque to 6.8 in-lb. Make sure the fold on
the Nomex shield is in the upper left corner of the OEM PCB.
c. Install the one remaining nut (161) and new 4-40
0.312 screw (252)
onto the OEM PCB; torque to 6.8 in-lb.
5. If the device is NOT equipped with the A06 OEM/SpO2 option, make sure
the boardstack shield is secured with two nuts (161) along the left side only;
torque to 6.8 in-lb.
6. Install the three remaining 4-40 nuts (161) and new 4-40
1.00 screw (125)
onto the A02 Patient Parameter PCB; torque to 6.8 in-lb.
7. Connect the W05 SpO2 Cable (if included) to the OEM PCB at J54 and
fasten with a screw (255).
Note:
Carefully align the SpO2 connector to the sockets, and gently press
the connectors into the sockets using steady pressure to avoid damage to
connector pins.
8. Connect the ECG cable to the A02 PP PCB at J23.
9. Complete the process by
.
9-87