LIFEPAK 20/20e Defibrillator/Monitor
Device Description
Functional Description
(continued)
W13 AC Power Cable
The W13 AC Power Cable connects the A09 AC Power Supply Module at J02
with the A03 Power Module (hard-wired connection) for the LIFEPAK 20
defibrillator/monitor, and at J49 for the LIFEPAK 20e defibrillator/monitor.
W14 Printer Flex Cable
The W14 Printer Flex Cable connects the A05 UI PCB at J34 with the A03 Power
Module at J45 and the A12 Printer.
W15 LCD to UI PCB
Cable
The W15 LCD to UI PCB Cable connects the A11 LCD Display PCB at CN1 with
the A05 UI PCB at J36.
W16 Display Jumper
Cable Extender
The W16 Display Jumper Cable Extender connects the A11 LCD Display PCB at
P77 to the A08 Backlight Inverter PCB at CN2.
W17 Backlight Inverter
Cable
The W17 Backlight Inverter Cable connects the A08 Backlight Inverter PCB at
P74 to the A05 UI PCB at J37.
W18 UI Flex Cable
The W18 UI Flex Cable connects the A02 Patient Parameters PCB at J21 and
J22 to the A05 UI PCB at J31.
3-40