LIFEPAK 20/20e Defibrillator/Monitor
Device Description
Functional Description
(continued)
W08 Battery Cable
On the LIFEPAK 20 defibrillator/monitor, the
W08 Battery Cable connects the
A07 Battery to the A03 Power Module. The cable is hardwired to the A03 Power
Module and the other end connects to the A07 Battery at J85.
On the LIFEPAK 20e defibrillator/monitor, the W08 Battery Cable connects the
A07 Battery at J85 to the A03 Power Module at J50.
W09/W10 Power to
Therapy PCB Cables
The W09 and W10 Power to Therapy PCB Cables connect the A03 Power
Module to the A04 Therapy PCB. W09 is a replaceable cable that connects to
the A04 Therapy PCB at J16 and to the A03 Power Module at J41.
On the LIFEPAK 20 defibrillator/monitor, W10 is hardwired to the A03 Power
Module and connects to the A04 Therapy PCB at J17.
On the LIFEPAK 20e defibrillator/monitor, W10 connects to the A04 Therapy
PCB at J17 and to the A03 Power Module at J51.
W11 ECG Sync/System
Cables
The W11 ECG Sync/System Cables connect the ECG sync connector and the
system connector to the A03 Power Module at J47.
W12 Grounding Cable
The
W12 Grounding Cable provides a grounding path for the Speed Dial.
3-39