LIFEPAK 20/20e Defibrillator/Monitor
Troubleshooting
Troubleshooting Chart
(continued)
Area
Observed Symptom
Suggested Corrective Action
ECG lead characteristics
(continued)
No amplitude ECG
Check or replace the ECG cable.
Replace the A02 Patient Parameter PCB
.
ECG gain out of tolerance
Check simulator output.
Check or replace the ECG cable.
Check or
replace the A02 Patient Parameter PCB
.
Therapy - Paddle ECG
lead characteristics
ECG gain out of tolerance
Check simulator output.
Check or replace the therapy cable.
Check or
.
ECG analog out (missing or
out of tolerance)
Check simulator output.
Check ECG on display.
Check W11 ECG out connector.
Remote sync
No remote sync
Turn on remote sync function.
Check ECG Out/Sync In connector. If bad,
.
Check or replace the W09 Power Cable.
Check or
.
(Continued on next page)
6-10