LIFEPAK 20/20e Defibrillator/Monitor
Replacement Procedures
Warnings and Cautions
The following general warnings and cautions apply to all actions you may
perform during maintenance of the device.
WARNING
SHOCK HAZARD. Servicing of this device must be performed by properly trained
individuals. This device may retain potentially lethal charges accessible inside the
device at any time, even when off. Follow the procedures carefully for discharging the
A13 Energy Capacitor.
The A13 Energy Capacitor carries high voltage. Discharge the capacitor before
handling.
It is possible to pinch and damage wires during reassembly. To avoid pinching wires,
carefully follow reassembly instructions.
CAUTION
Possible component damage.
The PCB assemblies contain static-sensitive
devices (SSDs). To avoid damage, observe the special handling practices
described under
Static-Sensitive Devices (SSD)
.
9-7