LIFEPAK 20/20e Defibrillator/Monitor
Troubleshooting
Service Code Table
(continued)
(Continued on next page)
Code
Service Code Description
Corrective Action
5032
Therapy USB disconnect
1. Clear error and perform PIP
2. Replace the A04 Therapy PCB
5033
PP USB download time out
1. Clear error and perform PIP
2. Reload the device software*
5036
USB initialization failed
1. Clear error and perform PIP
2. Reload the device software*
3. Replace the A01 System PCB
5037
USB unitization error
1. Clear error and perform PIP
2. Reload the device software*
3. Replace the A01 System PCB
5038
Cannot clear interrupt source
Replace the A01 System PCB
5039
Bus error reported by USB PP channel
1. Clear error and perform PIP
2. Replace the A02 PP PCB
3. Replace the A01 System PCB
6-25