LIFEPAK 20/20e Defibrillator/Monitor
Troubleshooting
Processing Service Codes
(continued)
Processing Service
Codes
(continued)
4. Complete the performance inspection procedure (PIP).
–
If completed successfully, the device may be returned to regular use.
(The service code(s) may have been related to EMI or ESD.)
–
If the service LED turns ON at any time during the PIP, stop the PIP and
continue to the next step in this procedure.
5. Consult the
for the suggested corrective action for
your PIP failure.
-OR-
Review the
codes, and then locate the service code in the
.
Note:
table to quickly jump to
the correct code in the Service Code Table.
6. Read the corrective action(s). If the corrective action calls for the
replacement of a part, click the link in the Troubleshooting Chart or click the
appropriate part in the footer at the bottom of the Service Code Table pages
to jump to the corrective action process.
7. Service the device based on these inputs, and then repeat the PIP.
8. For persistent error codes, contact your local Physio-Control service or sales
representative.
6-3