LIFEPAK 20/20e Defibrillator/Monitor
Replacement Procedures
Boardstack
(continued)
2. Install the new coin cell battery (344) by inserting battery (positive up)
underneath
the positive contacts and then pressing it into the negative side
of battery holder. Ensure that the battery is fullyseated in the holder.
3. After the battery has been replaced, check the voltage on the coin cell
battery holder circuit with a multi-meter to confirm that the battery is installed
correctly and is making contact with the holder.
4. Replace the PCB shield (127).
5.
Install the A02 Patient Parameter PCB and A06 OEM PCB
9-93