LIFEPAK 20/20e Defibrillator/Monitor
Replacement Procedures
Boardstack
(continued)
A02 Patient Parameter
and A06 OEM/SpO2
Assembly Installation
To install the A02 Patient Parameter and optional A06 OEM/SpO2 assembly:
1. Make sure the orange parameter ISO mount (258) is installed on the A02
Patient Parameter PCB with the square end facing out.
2. If the device is equipped with the A06 OEM/SpO2 option:
a. Make sure the three, white, OEM ISO mounts (149) are installed on the
A06 OEM PCB with the square ends facing out.
b. Connect the A02 Patient Parameter to the A06 OEM/SpO2 assembly at
J24.
3. Install the A02 Patient Parameter (OEM/SpO2) PCB onto the five standoffs
(129) on the boardstack shield (ensure the standoffs are tight and in good
condition). Make sure the PCB clears the lip in the lower right corner and the
60-pin connector seats correctly.
4. If the device is equipped with the A06 OEM/SpO2 option:
(Continued on next page)
9-86