LIFEPAK 20/20e Defibrillator/Monitor
Replacement Procedures
Boardstack
(continued)
A02 Patient Parameter
and A06 OEM/SpO2
Assembly Removal
(continued)
6. Inspect the orange parameter ISO mount (258) installed in the A02 Patient
Parameter PCB. Verify the mount is in good condition.
7. If the device is equipped with the A06 OEM/SpO2 option:
a. Inspect the three, white, OEM ISO mounts (149) installed in the A06
OEM PCB. Verify the mounts are in good condition.
b. Separate the A02 Patient Parameter PCB from the A06 OEM/SpO2
assembly at J24.
Note:
If replacing the A02 Patient Parameter PCB or A06 OEM PCB, remove
the isolated mounts from the old PCBs, note the condition, and install
them on the new PCBs. Replace the isolated mounts if broken or
cracked.
258
9-85