414
C
o
nt
rol
w
it
h C
om
m
uni
c
a
tion P
or
t
C
h
ap
te
r 9
Connector
(Binary Window)
Parameter
Read Write Minimum
value
Maximum
value
Details
Common Judge. Max.
CBWnnn:CMAX
Yes
Yes
0
62500
Common Judge. Min.
CBWnnn:CMIN
Yes
Yes
0
62500
Common Judge. Max.
and Min.
CBWnnn:CPAIR
Yes
Yes
0
62500
Data 1 = Min.
Data 2 = Max.
Individual Judge. Max.
*1)
CBWnnn:MAX.ss
Yes
Yes
0
62500
Individual Judge. Min.
*1)
CBWnnn:MIN.ss
Yes
Yes
0
62500
Individual Judge. Max.
and Min. *1)
CBWnnn:LPAIR.ss
Yes
Yes
0
62500
Data 1 = Min.
Data 2 = Max.
Connector
(Gray Window)
Parameter
Read Write Minimum
value
Maximum
value
Details
Common Judge. Max.
CGWnnn:CMAX
Yes
Yes
0
255
Common Judge. Min.
CGWnnn:CMIN
Yes
Yes
0
255
Common Judge. Max.
and Min.
CGWnnn:CPAIR
Yes
Yes
0
255
Data 1 = Min.
Data 2 = Max.
Individual Judge. Max.
*1)
CGWnnn:MAX.ss
Yes
Yes
0
255
Individual Judge. Min.
*1)
CGWnnn:MIN.ss
Yes
Yes
0
255
Individual Judge. Max.
and Min. *1)
CGWnnn:LPAIR.ss
Yes
Yes
0
255
Data 1 = Min.
Data 2 = Max.
Connector
(Gray Edge)
Parameter
Read Write Minimum
value
Maximum
value
Details
Max. Judge. Pitch
CGEnnn:PMAX
Yes
Yes
0
99999999.999
Min. Judge. Pitch
CGEnnn:PMIN
Yes
Yes
0
99999999.999
Max. and Min.
Judgement Pitch
CGEnnn:PPAIR
Yes
Yes
0
99999999.999 Data 1 = Min.
Data 2 = Max.
Max. Distance
Difference Judgement
CGEnnn:UMAX
Yes
Yes
0
99999999.999
Min. Distance
Difference Judgement
CGEnnn:UMIN
Yes
Yes
0
99999999.999
Max. and Min.
Distance Difference
Judgement
CGEnnn:UPAIR
Yes
Yes
0
99999999.999 Data 1 = Min.
Data 2 = Max.
Smart Edge (Circle)
Parameter
Read Write Minimum
value
Maximum
value
Details
Max. Measurement *2) SECnnn:LMAX
Yes
Yes
0
99999999.999
Min. Measurement *2)
SECnnn:LMIN
Yes
Yes
0
99999999.999
Max. and Min.
Measurement *2)
SECnnn:LPAIR
Yes
Yes
0
99999999.999 Data 1 = Min.
Data 2 = Max.
Max. Standard
Deviation
SECnnn:SMAX
Yes
Yes
0
99999999.999
Min. Standard
Deviation
SECnnn:SMIN
Yes
Yes
0
99999999.999
Max. and Min.
Standard Deviation
SECnnn:SPAIR
Yes
Yes
0
99999999.999 Data 1 = Min.
Data 2 = Max.
Max. Circularity
SECnnn:CMAX
Yes
Yes
0
99999999.999
Min. Circularity
SECnnn:CMIN
Yes
Yes
0
99999999.999
Max. and Min.
Circularity
SECnnn:CPAIR
Yes
Yes
0
99999999.999 Data 1 = Min.
Data 2 = Max.
Max. Unused Edge
Count
SECnnn:NMAX
Yes
Yes
0
3597
*1) "ss" in the parameters of Individual Judgement is cell numbers.
*2) "Measurement": Radius, Width or Diameter (Determined by the selection of "Detect Mode".)
Содержание PV-500
Страница 1: ......
Страница 10: ......
Страница 11: ...9 Names and Descriptions Chapter 1 INDEX Chapter 1 Names and Descriptions ...
Страница 21: ...19 Installation and Wiring Chapter 2 INDEX Chapter 2 Installation and Wiring ...
Страница 37: ...35 Basic Operation Chapter 3 INDEX Chapter 3 Basic Operation ...
Страница 55: ...53 Setting Inspection Conditions Chapter 4 INDEX Chapter 4 Setting Inspection Conditions ...
Страница 315: ...313 Setting of Operation and RUN Menu Chapter 5 INDEX Chapter 5 Setting of Operation and RUN Menu ...
Страница 331: ...329 Environment Settings Chapter 6 INDEX Chapter 6 Environment Settings Changing System Settings ...
Страница 345: ...343 Chapter 7 Changing Configuration INDEX Chapter 7 Changing Configuration TOOL ...
Страница 375: ...373 Operation and Data Output through I O Terminals Chapter 8 INDEX ...
Страница 390: ......
Страница 391: ...389 Control with Communication Port Chapter 9 INDEX Chapter 9 Control with Communication Port ...
Страница 430: ......
Страница 431: ...429 PVWIN Chapter 10 INDEX Chapter 10 PVWIN ...
Страница 437: ...435 Chapter 11 Specifications and List of Product Numbers INDEX Chapter 11 Specifications and List of Product Numbers ...
Страница 493: ......
Страница 494: ......