59
2-4-2 Connecting ground and power supply
This section describes how to connect ground and power supply. For the AC power cable, use the attached AC power
cable. The power terminal, protective conductor terminal and frame ground terminal are located on the back side of the unit
as shown below.
Power
terminal
Power
terminal
Controller
LP-M200/M205/M500/M505
LP-MA00/MA05/MA06
Head
Protective conductor
terminal
Protective conductor
terminal
Frame ground
terminal: F.G.
WARNING
• Perform the connection of the Power Terminal Block with power-off state.
There is a risk of electrical shock.
1.
Connect the protective conductor terminal and the frame
ground terminal to the grounding part.
Screw size: M4
Tightening torque: 1.0N·m
ワㄐㄕㄊㄆ
• Before connecting the power supply, connect the protective conductor terminal and the frame ground terminal
permanently to the grounding part.
• Install such that the controller housing and the head housing are at the same electric potential.
2.
Loosen two M3 screws of the power terminal casing and remove
it.
3.
Remove the resin cover on the terminal.
Protective conductor terminal /
Frame ground terminal
Ground
Terminal
Resin cover
ME-LPM-SSM-10
Содержание LP-M Series
Страница 16: ...16 MEMO ME LPM SSM 10...
Страница 27: ...27 Chapter 1 Specification ME LPM SSM 10...
Страница 37: ...37 Chapter 2 Preparation ME LPM SSM 10...
Страница 69: ...Chapter 3 Connection for External Control ME LPM SSM 10...
Страница 128: ...Chapter 4 Maintenance ME LPM SSM 10...
Страница 148: ...Troubleshooting ME LPM SSM 10...
Страница 169: ...Index ME LPM SSM 10...
Страница 172: ...172 MEMO No 9000 0066 34V ME LPM SSM 10...
Страница 173: ...No 9000 0066 34V...