User Manual
– Technical characteristics
NIMXEN020I
Owandy Radiology SAS
26
Tube-head characteristics
Model
MPV 05
Manufacturer
Owandy Radiology
Maximum tube voltage
86 kV
kV accuracy
± 8 %
Maximum anodic current
12.5 mA
Anodic current accuracy
± 10 %
Duty cycle
1 : 16
Reference loading conditions related to
maximum energy input to the anode
2812.5mAs/h @86 kV
Nominal power
1.075 kW (86 kV - 12.5 mA)
Total filtration
≥ 2.5 mm Al eq. @ 86 kV
HVL (Half value layer)
> 3.2 mm Al eq. @ 86 kV
Transformer insulation
Oil bath
Target angle and reference axis
See Figure 2
Cooling
By convection
Leakage radiation at 1 m
< 0.5 mGy/h @ 86 kV - 12.5 mA - 3s duty cycle
1/16
Tube-head maximum thermal capacity
310kJ
Figure 2: Tube-head target angle (view from the bottom)