User Manual
– Making an exam
NIMXEN020I
Owandy Radiology SAS
71
10.
MAKING AN EXAM
Demo video: https://www.youtube.com/watch?v=0_ctfc1lr-0
Note
With paediatric patient it is recommended to have a greater attention in taking exams.
In general with child and adolescent patients, before taking the exposure it is suggested
to run the exam in test mode in such a way to show the young patient how the unit
works and make them feel comfortable.
To make an exam, proceed as follows:
1. Press the power switch located on the upper part of the equipment on the operator side to
position "1". This will start the "CHECK" function, which is indicated by the LEDs lighting up.
When the "CHECK" function is complete, the green LED on the equipment keyboard starts
blinking.
2. Press >O< button on the keyboard to run the equipment axis zero.
Warning
During equipment axis zero reset, check that the unit does not collide with external
object.
3. Run the Virtual Interface on the PC and wait for the connection between the PC and
equipment; this status is indicated by the blue LED lighting up on the keyboard and on the
Virtual Interface.
4. Select the exam and parameters on the Virtual Interface.
5. Press the button ">O<" on the keyboard to run the chin rest automatic positioning accordingly
with the current exam selection.
6. The white LED light ON, fixed in Panoramic, Bitewing and 3D or blinking slowly in other
exams.
7. Place the proper chin rest (see paragraph 9.2) or the dental model/dental impression support
corresponding to the current exam selection.
8. If the chin rest or support is correct, the white LED is steady on, otherwise it will blink quickly.
Note
To go back in the exam selection status to change exam or settings, the proper chin
rest of the current exam has to be placed in position, then press >O< to go in start exam
position and then, when green led is ON, press >O< again to go back in exam selection
status.
9. Position the patient with the help of the lasers then close the temple clamps.
10. Press >O< button to put the equipment in the start exam position; the green LED lights up:
the unit is now ready for X-rays.
Warning
Since the X-ray button is a "dead man's switch", its release before the end of the
exposure, immediately stops the X-ray emission and the arm rotation.
Error 362 or Error 760 will be displayed.