User Manual
– Quality assurance program
NIMXEN020I
Owandy Radiology SAS
44
8.2
Functioning of the indicator lights
Power ON the unit, verify that the "Machine Ready" (1), "X-Ray Emission" (2) and "Computer
connection" (3) LEDs blink twice.
Figure 8
In case the test fails, verify that the main power supply is present in the room. If the case, call
technical assistance.
8.3
Laser alignment check
Power ON the unit and perform the axis reset by pressing the >O< button.
At the end of the axis positioning, select standard Panoramic exam (see paragraph 12.1) and
press >O<. Place the support plate (Figure 10) on the chin rest support and power ON the laser.
Check that the mid-sagittal laser beam is aligned to the reference line of the support plate (±
3mm).
At the end of the check, switch OFF the unit.
Laser correctly aligned
Laser misaligned
Figure 9
In case the test fails, repeat it checking that there is no mechanical interference. If misalignment
is still present, call technical assistance.