User Manual
– Technical characteristics
NIMXEN020I
Owandy Radiology SAS
25
Image magnification
Geometric
magnification
Magnification after
software correction
Adult / Child standard Panoramic
1 : 1.28 (constant over
dentition part)
1 : 1 (*)
TMJ open/closed mouth
1 : 1.25 (nominal)
1 : 1 (*)
Sinus
1 : 1.27 (nominal)
1 : 1 (*)
(*) Warning
The declared image magnification value is valid after proper software calibration.
Note
I-MAX 3D is based on a standard dentition and ascending rami shape.
This shape, based on statistical studies, establishes a form for the dentomaxillofacial
complex, adopted as "standard".
I-MAX 3D follows a rototranslation path which maintains the magnification factor as
stated in the Technical Characteristics of each type of exam as constant along this
"standard" shape only along the dentition area. The patient’s anatomy can differ
significantly from the statistical model, so the magnification factor is not maintained
and may be different from the value stated. Based on experience and competence,
the user has to judge this variation.
IN ANY CASE, RADIOGRAPHY IMAGES CANNOT BE USED TO PERFORM
CALCULATIONS OF DISTANCES, ANGLES ETC. ON THE IMAGE.