User Manual
– Quality assurance program
NIMXEN020I
Owandy Radiology SAS
58
6.
Take a second exposure setting the following parameters: 86 kV, 12.5 mA, 3 s and verify
that the measured values are in the acceptance limits listed in the following table.
kV
mA
t (s)
kV acceptance limits
Time acceptance limits
60
2
3
55.2 to 64.8 kV
2.85 to 3.15 s
86
12.5
3
79.1 to 92.8 kV
2.85 to 3.15 s
7.
In case the test fails (result do not match the indicated values), proceed with the following
actions:
•
Check the probe position and repeat the test
•
If the values are still out of range, perform the test using the invasive method as
described in the Service Manual
•
If the values are still out of range, call technical assistance.
8.
Record the test results in the log book at paragraph 7.6.1.