User Manual
– 3D exam
NIMXEN020I
Owandy Radiology SAS
85
13.5 3D Maxillary Teeth
The acquired volume is a 50 x 50 mm FOV centered on five
different positions along the maxillary arch, depending on
the user selection.
Two acquisition modes are available: Standard definition
(175 µm voxel size) for a standard resolution 3D acquisition,
and High definition (87.5 µm voxel size) for the enhancement
of the finest details.
13.6 3D Mandibular Teeth
The acquired volume is a 50 x 50 mm FOV centered on five
different positions along the mandibular arch, depending on
the user selection.
Two acquisition modes are available: Standard definition
(175 µm voxel size) for a standard resolution 3D acquisition,
and High definition (87.5 µm voxel size) for the enhancement
of the finest details.
13.7 3D Extended Volume / 3D Airways
The acquired volume is a 116x103 mm FOV either centered to
the patient dentition or back shifted towards the upper respiratory
tract.
In case the collimator kit code 6604061200 is present, these
exams are not available.
Warning
The portion of image volume in the "extended" area is reconstructed using a partial set
of projections.
Compared to the central part of the volume, such image portion can have lower
definition in the anatomical details.