6.6
CBCT Conditions of Operation
The following table lists the conditions of operation for the unit working in CBCT modality.
Quantity
Range
Tube current (mA)
From 2 to 12.5 mA
kV
From 60 to 86 kV
Exposure time
21.2 s
X-ray filtration
≥ 2.5 mm Al eq. @ 86 kV
Nominal Tomographic section thickness
0.175 mm (0.0875 mm in HR)
Image receptor are
144 x 118.6 mm
6.6.1
Reference Plane
The reference plane offset is the horizontal plane passing on the chin rest of the unit.
The Figure 6 shows the position of the reference plane and its location with respect to the chin
rest, the focal spot and the irradiated volume by the X-ray cone beam. Each exam has a proper
chin support that gives the proper reference plane offset.