User Manual
– Making an exam
NIMXEN020I
Owandy Radiology SAS
72
Note
The rotation of the arm and the emission of the X-rays will start with a delay of about 3
seconds from when the X-ray button is pressed.
Note
In this state kV and mA can be adjusted, while patient size, patient type and biting can't
be changed.
Note
Ready for X-ray status is signalled by the green LED on the equipment (3 - Figure 17)
and on the GUI lighting up.
Ready for X-ray status remains as long as the equipment is in the start exam position
and the GUI is connected to the equipment.
11. Press the X-ray button for the entire duration of the exposure.
Note
When the "Test" key is selected, the Test function is activated. In this condition, it is
possible to make the unit carry out all movements made during the exam without
emitting X-rays.
Test mode may also be useful when the equipment is used on child patients, to show
them how the equipment works before running the exam.
Once the cycle is completed, deactivate the "Test" function by pressing the key again.
12. Once the exposure is completed, the system will rotate back to patient exit position. It is now
possible to free the patient from the positioning device.
13. Press >O< to return to axis 0 position.
Warning
In case of exam interruption, let the patient exit the unit, follow the instruction on the
GUI then press >O< to carry out the axis reset.
Warning
During the emission of X-rays, protection procedures for the operator and personnel in
the area must comply with local regulations. In all cases, it is recommended that only
the patient and operator are present in the room during the emission of X-rays. If the
operator is not protected by suitable screens, he must stand at least 2 meters away
from the emission of the rays (Figure 1).
Note
I-MAX 3D assumes that the digital sensor is ready: if this is not the case, the blue light
indicator of "Computer connection" status (5 - Figure 17) start blinking slowly.
To reset the message on I-MAX 3D, press "OK" on the GUI and follow the instruction
provided (if on the equipment keyboard the green light indicator of "Machine Ready"
status is blinking, press >O< to perform the axis reset).