– General instruction for use
NIMXEN020I
Owandy Radiology SAS
61
The unit is now ready for exams. To run the exam, press the X-ray button and keep it pressed
until the exam end.
Warning
Since the X-ray button is a "dead man's switch", its release before the end of the
exposure, immediately stops the X-ray emission and the arm rotation.
Error 362 or Error 760 will be displayed.
Note
In the start exam position, the laser light and column movement are not enabled, and
on the GUI only kV and mA adjustment is allowed.
Note
Ready for X-ray status is signalled by the green LED on the equipment (3 - Figure 17)
and on the GUI lighting up.
Ready for X-ray status remains as long as the equipment is in the start exam position
and the GUI is connected to the equipment.
To switch OFF the unit press the power switch located on the upper part of the equipment on the
operator side to position "0".
The LEDs will go off.
9.1.1
Emergency button
The equipment has a red emergency button located on the upper part of the unit, near the power
switch.
The emergency button only stops the vertical column movement.
In case of an emergency column situation, press the emergency button to stop the movement.
If columns don't move, check that the emergency button is not pressed; rotate the button to
release it.