Ossila Slot-Die Coater
User Manual
Ossila.com
33
Ossila Limited © 2019
1.
Once the required parameters have been
set in
Program Mode
, and both the slot-die
head and syringe have been installed (see
Sections 9.2.5
and
9.2.7
respectively), the
program can be run by pressing the
button. A green progress bar will appear
showing the overall progress.
Note: If the stage delay is set in the
program, the stage delay icon will appear
blinking next to the elapsed time until delay
time is up.
Note: If the coating speed has not been set
for all off the program steps, a warning will
appear to notify the user. Running will be
aborted until the speed is set.
2.
You can start coating at any stage position
by positioning the stage in
Manual Mode
.
However, the position is tracked so that it
will not exceed the maximum travel
distance of 100 mm. Where the total
coating length plus the initial stage position
exceeds this maximum travel distance, a
warning will appear. The stage position
should be put closer to the default home
position (see
Section 9.2.1
), or the coating
length should be reduced.
Stage Delay
Icon