Ossila Slot-Die Coater
User Manual
Ossila.com
20
Ossila Limited © 2019
4.
The surface temperature of the hotplate
may differ slightly from the displayed
reading. To set the exact temperature value
on the surface, a temperature offset can be
applied
by navigating to the ‘Temp offset’
line and pressing
to edit. You can set
the temperature offset by
±10 °C.
To
change the sign for ‘Temp offset’, navigate
to the sign and press
or
to change
it.
Note: An external temperature sensor is
required to measure the surface
temperature (not supplied).
Note: The displayed temperature has the
offset applied and therefore will appear
higher than the set point temperature when
a positive offset is applied.
5.
To change the syringe size, navigate to the
‘
Syringe size
’
line and press
to enter
edit mode. To change to different sizes of
the syringe, press
or
to scroll
between the pre-set sizes. Press
again
to exit the edit mode.
Note: The selected syringe size is displayed
in the upper right corner within
Program
Mode
.
6.
A custom syringe size can be entered by
selecting ‘Cust’, which allows the
syringe
diameter to be manually entered.
Note: If the diameter value is not set, a
warning will appear on the screen to notify
the user.