Ossila Slot-Die Coater
User Manual
Ossila.com
13
Ossila Limited © 2019
6.4
Emergency Stop
The slot-die coater is fitted with an emergency stop button that can be pressed in the event of
emergencies e.g. the trapping of objects within pinch points. When pressed, the emergency stop
button cuts off the power to the system and a buzzer will turn on to indicate that the button has
been pressed. The emergency stop button will remain pressed until the button is rotated clockwise
and released. Once released, the system can be restarted by turning the unit off and on again using
the on/off switch at the back of the unit. The emergency stop button should not be used for
powering down the system after use and should only be used in emergency situations.
Figure 6.1
The emergency stop button mounted to the front of the slot-die coater.
6.5
Tube Clip
A tube clip has been provided with the system. The clip chain attaches to the slot-die head using a
simple washer. The clip itself can then be attached to any PTFE tubing that is connected to the
system. This ensures that if the polypropylene connector becomes damaged during operation, the
tubing is held in place and solution does not escape the system. The clip should be attached
whenever the system is used.
Figure 6.2
The slot-die coater tube clip and chain is used to secure the tubing