Ossila Slot-Die Coater
User Manual
Ossila.com
29
Ossila Limited © 2019
3.
Level the stage across the length
•
Using the
button, move the stage
so that the gauge rollers are now
situated at the front end of the stage.
•
Using the front thumbscrew, adjust
the height of the front end of the
stage until the digital height gauges
read close to zero on both sides.
•
The difference between either side
should be no more than 50 µm.
4.
The stage should already be at or close to the stage home position. To ensure this is the case
return the stage to the stage home position by pressing the
button. To increase the
accuracy of the levelling, repeat steps 2 and 3.
9.2.7
Installing the Syringe
1.
There are two sets of holes on the plunger
pusher to accommodate different sized
syringes. The height of the plunger clip can
be adjusted by changing which set of holes
the securing screws pass through. The
upper pair of holes can be used for large
syringes, and the lower pair for small
syringes.
2.
Fill your syringe with the desired quantity of
solution. Ensure air bubbles are expelled
from the syringe.