Ossila Slot-Die Coater
User Manual
Ossila.com
28
Ossila Limited © 2019
9.2.6
Levelling the Stage
PLEASE NOTE: The digital height gauges used to level the stage have a resolution of 1 µm. It is not
possible to level the stage to this accuracy. Recommended levelling parameters are given in the
procedure below. If substrate heating is required, the hotplate temperature should be set, the
hotplate turned on and the stage allowed to reach the set point temperature
before
levelling the
stage.
1.
Zero height gauges relative to the head:
•
In
Manual Mode
, use the
button,
move the stage so that the gauge
rollers are no longer touching the top
surface.
•
Turn the digital height gauges on by
pressing ‘OFF/ON’ and set them to zero
by pressing ‘Zero’.
2.
Level the stage across the width:
•
Using the
button, move the stage
so that the gauge rollers are on the
top surface of the stage.
•
Using the left and right adjustment
thumbscrews at the back of the stage,
level the digital height gauges so that
the readings on both dials are within 50
µm of each other.
Press the ‘
Z
ero’
button on both digital height gauges so
they both read zero.