Adjusting the analog output value
Discrepancies may occur between the current value/voltage value output as analog set on the Sensor
Controller and the current value/voltage value actually measured due to the conditions for the connected
external device or other factors.
The analog output adjustment function can be used to correct this discrepancy.
The output values are corrected by entering the adjustment value for the current or voltage values for any two
points.
Set the output destination and select either current or voltage output beforehand. Also, connect the analog output
signal line to an external ammeter or voltmeter.
1
Set the operating mode to the FUN mode.
3-3 Switching operation modes p.61
2
Select ON/OFF from [Monitor Focus].
3
Enter the [Distance] and [Output value] at
[Point1].
4
Likewise, enter the [Distance] and [Output
value] at [Point2].
Item
Setting item Setting value
Description
Analog out-
put adjust-
ment
Analog output adjustment
ON/OFF (default value)
Set analog output correction ON/OFF.
Point1
Reference value
(current/value)
4 to 20 [mA]/-10 to 10 [V]
Sets the current or voltage to be used as the correction
reference in the entry field on the left.
adjustment value -999 to 999
Sets the adjustment value when the reference value is
measured in the entry field on the right.
Point2
Reference value
(current/value)
4 to 20 [mA]/-10 to 10 [V]
Sets the current or voltage to be used as the correction
reference in the entry field on the left.
adjustment value -999 to 999
Sets the adjustment value when the reference value is
measured in the entry field on the right.
1
Set the operating mode to the FUN mode.
3-3 Switching operation modes p.61
Important
Содержание ZW-C1*AT
Страница 17: ...ZW User s Manual 1 15 ...
Страница 19: ...1 Basic configuration Basic configuration 1 1 ZW series Displacement Sensors 18 1 2 Basic Operation Flow 19 ...
Страница 54: ...Installing the Sysmac Studio 52 ZW User s Manual MEMO ...
Страница 74: ...Operating with Sensor Controller 72 ZW User s Manual MEMO ...
Страница 120: ...Setting the System 118 ZW User s Manual MEMO ...
Страница 138: ...Controll input signal with PC tool 136 ZW User s Manual MEMO ...
Страница 262: ...No protocol Connection 260 ZW User s Manual MEMO ...
Страница 270: ...Offline Debugging of the Sensor Control Program and Sensor Operation 268 ZW User s Manual MEMO ...
Страница 271: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 270 8 2 Troubleshooting 284 ...
Страница 358: ...Connecting by No protocol Communications 356 ZW User s Manual MEMO ...
Страница 449: ......