
Displaying measured values in graphs
124
ZW User's Manual
Starting and ending sampling before and after the trigger condition is met
You can adjust how long to start or end the sampling before or after the condition for starting and ending
sampling is met.
• Sampling start position A
To start sampling before the time when the trigger start condition is met, input a negative value for the delay
time.
• Sampling start position B
To end sampling after the time when the trigger end condition is met, input a positive value for the delay time.
1
Input the delay time.
Item
Setting item
Range
Description
Trigger condition Delay time
-999.999999 to
999.999999 ms
The sampling starts or ends at the time shifted by the set
time from when the trigger condition is met.
Measurement data satisfied the start condition trigger
Measurement value
Sampling interval
Sampling start position A
Sampling start position B
Содержание ZW-C1*AT
Страница 17: ...ZW User s Manual 1 15 ...
Страница 19: ...1 Basic configuration Basic configuration 1 1 ZW series Displacement Sensors 18 1 2 Basic Operation Flow 19 ...
Страница 54: ...Installing the Sysmac Studio 52 ZW User s Manual MEMO ...
Страница 74: ...Operating with Sensor Controller 72 ZW User s Manual MEMO ...
Страница 120: ...Setting the System 118 ZW User s Manual MEMO ...
Страница 138: ...Controll input signal with PC tool 136 ZW User s Manual MEMO ...
Страница 262: ...No protocol Connection 260 ZW User s Manual MEMO ...
Страница 270: ...Offline Debugging of the Sensor Control Program and Sensor Operation 268 ZW User s Manual MEMO ...
Страница 271: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 270 8 2 Troubleshooting 284 ...
Страница 358: ...Connecting by No protocol Communications 356 ZW User s Manual MEMO ...
Страница 449: ......