Displaying measured values in graphs
122
ZW User's Manual
Specifying the sampling start and end conditions
You can specify the conditions for starting and ending sampling.
1
Check the Trigger start conditions/Trigger end conditions checkbox.
2
Select the trigger condition.
3
Select the trigger target.
Item
Setting item
Range
Description
Sampling setting
Trigger start
conditions
Input and output
Specify parallel I/O (TIMING, ZERO, BUSY, ENABLE,
HIGH, PASS, LOW) as the trigger condition.
Data slope
Sets the change in the measured values for TASK1-4 as
the start condition. Sets the trigger level. When the
measured value rises above this value or falls below it, the
start trigger is issued.
Data window (In)
Sets the change in the measured values for TASK1-4 as
the start condition. When the measured value is inside the
range, the start trigger is issued.
Data window (Out)
Sets the change in the measured values for TASK1-4 as
the start condition. When the measured value is outside
the range, the start trigger is issued.
Trigger end
conditions
Input and output
Specify parallel I/O (TIMING, ZERO, BUSY, ENABLE,
HIGH, PASS, LOW) as the trigger condition.
Data slope
Sets the change in the measured values for TASK1-4 as
the start condition. Sets the trigger level. When the
measured value rises above this value or falls below it, the
end trigger is issued.
Data window (In)
Sets the change in the measured values for TASK1-4 as
the start condition. When the measured value is in the
range, the end trigger is issued.
Data window (Out)
Sets the change in the measured values for TASK1-4 as
the start condition. When the measured value is outside
the range, the end trigger is issued.
Number of data
The sampling data is counted from when the start
condition is met and when the specified number of data
points have been sampled, the end trigger is issued.
Item
Setting item
Range
Description
Trigger target
I/O
TIMING
TIMING input signal (parallel I/O)
ZERO
ZERO input signal (parallel I/O)
BUSY
BUSY output signal (parallel I/O)
ENABLE
ENABLE output signal (parallel I/O)
HIGH
HIGH output signal (parallel I/O)
PASS
PASS output signal (parallel I/O)
LOW
LOW output signal (parallel I/O)
Data slope
Data window (in)
Data window (out)
TASK1 TASK1
measurement
results
TASK2
TASK2 measurement results
TASK3
TASK3 measurement results
TASK4
TASK4 measurement results
Содержание ZW-C1*AT
Страница 17: ...ZW User s Manual 1 15 ...
Страница 19: ...1 Basic configuration Basic configuration 1 1 ZW series Displacement Sensors 18 1 2 Basic Operation Flow 19 ...
Страница 54: ...Installing the Sysmac Studio 52 ZW User s Manual MEMO ...
Страница 74: ...Operating with Sensor Controller 72 ZW User s Manual MEMO ...
Страница 120: ...Setting the System 118 ZW User s Manual MEMO ...
Страница 138: ...Controll input signal with PC tool 136 ZW User s Manual MEMO ...
Страница 262: ...No protocol Connection 260 ZW User s Manual MEMO ...
Страница 270: ...Offline Debugging of the Sensor Control Program and Sensor Operation 268 ZW User s Manual MEMO ...
Страница 271: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 270 8 2 Troubleshooting 284 ...
Страница 358: ...Connecting by No protocol Communications 356 ZW User s Manual MEMO ...
Страница 449: ......