79
Z500
Operation Manual
SECTI
ON 4
SECTION 4
Menus for Expert Menu
Height
in Example 1 is taken here to describe the setting method.
Make the necessary adjustments when setting a different item.
1.
Move the cursor to
Height
and press the
ENT
Key.
The selections will be displayed.
2.
Select
Calibrate
.
Calibrate to 5 mm.
5mm
5mm
4.95mm
Measurement of point 1
Measurement of point 2
Movement distance from the edge position: 5 mm
Use the method shown below to calibrate workpieces so that a displacement value of 4.95 mm is
measured as 5 mm.
Input the
position
of point 2.
5mm
Input the
position
of point 1.
0mm
Compensate for this error.
Measurement value
Workpiece Movement
Measure with the workpiece
at the reference position.
Example 2
: When selecting
Edge position
Move the workpiece by 5 mm
for the measurement.
OUT0:Height
OUT1:Step: 2 pts
OUT2:Edge position
OUT3:Define
OUT4:
OUT5:
OUT6:
OUT7:
Scn 0 Set
Change
Calibrate
Output
Delete
Scn 0 Set
St
e
p
2
St
e
p
1
St
e
p
3
St
e
p
4
St
e
p
5
St
e
p
6
St
e
p
7
Содержание Z500 Series
Страница 20: ...18 SECTION 1 1 2 Operational Flow Z500 Operation Manual SECTION 1 Features MEMO ...
Страница 40: ...38 SECTION 2 2 4 Saving Settings and Exiting the Z500 Z500 Operation Manual SECTION 2 Basic Operations MEMO ...
Страница 144: ...142 SECTION 5 5 8 Backing Up Data to a Computer Backup Z500 Operation Manual SECTION 5 Other Functions MEMO ...
Страница 184: ...182 7 3 Analog Output SECTION 7 Z500 Operation Manual SECTION 7 I O Format MEMO ...