Z500
Operation
Manual
7-2 R
S
-232C
SE
CTION 7
173
SECTION 7
I/O Format
• TREND (or T)
Function
……
Obtains the measurement values from the start to the end of the
measurement under the Trigger setting.
CHECK
Perform the setting as shown below, before using the command.
Set the trigger input method to
External
or
Self
.
REFERENCE
For details on the Trigger function, refer to page 90.
REFERENCE
For details on how to set the Trend Monitor's display contents, refer to page 118.
TREND (or T) <Output number> <Output coefficient>
• Specify an output number between 0 and 7.
• Specify an output coefficient between 1 and 1,000.
• The output coefficient is the number of intervals between measurements that
are output.
• The number of data items that will be output can be calculated using the
formula below.
Measurement time
÷
Sampling time = Number of measurements
Number of measurements
÷
Measurement coefficient = Number of data items
output (The number is rounded up to the nearest integer.)
Output number
Meaning
0
OUT0
1
OUT1
2
OUT2
3
OUT3
4
OUT4
5
OUT5
6
OUT6
7
OUT7
Input
delimiter
XX.XXms
Thru
Scn 0 Run
Sampling time
Measurement results
Sampling time
Start to End (Measurement time)
(Selectable in Trigger setting.)
Содержание Z500 Series
Страница 20: ...18 SECTION 1 1 2 Operational Flow Z500 Operation Manual SECTION 1 Features MEMO ...
Страница 40: ...38 SECTION 2 2 4 Saving Settings and Exiting the Z500 Z500 Operation Manual SECTION 2 Basic Operations MEMO ...
Страница 144: ...142 SECTION 5 5 8 Backing Up Data to a Computer Backup Z500 Operation Manual SECTION 5 Other Functions MEMO ...
Страница 184: ...182 7 3 Analog Output SECTION 7 Z500 Operation Manual SECTION 7 I O Format MEMO ...